Size: | Price: |
---|---|
48 T | $320.00 |
96 T | $458.00 |
Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
10.00 | 2.255 | 2.147 |
5.00 | 1.636 | 1.528 |
2.50 | 1.226 | 1.118 |
1.25 | 0.897 | 0.789 |
0.63 | 0.475 | 0.367 |
0.32 | 0.312 | 0.204 |
0.16 | 0.302 | 0.194 |
0.00 | 0.108 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 80-97% | 88% |
EDTA plasma(n=5) | 77-95% | 86% |
Heparin plasma(n=5) | 94-106% | 100% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 86-101% | 86-101% | 87-96% | 92-105% |
EDTA plasma(n=5) | 89-102% | 92-101% | 88-102% | 87-102% |
Heparin plasma(n=5) | 85-96% | 87-101% | 89-102% | 85-98% |