Size: | Price: |
---|---|
48 T | $320.00 |
96 T | $458.00 |
Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
10.00 | 2.054 | 1.981 |
5.00 | 1.528 | 1.455 |
2.50 | 1.194 | 1.121 |
1.25 | 0.865 | 0.792 |
0.63 | 0.614 | 0.541 |
0.32 | 0.324 | 0.251 |
0.16 | 0.179 | 0.106 |
0.00 | 0.073 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 87-95% | 91% |
EDTA plasma(n=5) | 93-107% | 100% |
Heparin plasma(n=5) | 87-99% | 93% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 85-92% | 97-103% | 87-98% | 95-102% |
EDTA plasma(n=5) | 98-103% | 86-92% | 87-93% | 93-102% |
Heparin plasma(n=5) | 85-94% | 97-106% | 92-101% | 82-90% |