Size: | Price: |
---|---|
48 T | $320.00 |
96 T | $458.00 |
Concentration (pg/mL) | OD | Corrected OD |
---|---|---|
3000.00 | 1.952 | 1.868 |
1500.00 | 1.562 | 1.478 |
750.00 | 1.174 | 1.090 |
375.00 | 0.883 | 0.799 |
187.50 | 0.567 | 0.483 |
93.75 | 0.324 | 0.240 |
46.88 | 0.213 | 0.129 |
0.00 | 0.084 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 90-105% | 97% |
EDTA plasma(n=5) | 87-99% | 93% |
Heparin plasma(n=5) | 82-95% | 88% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 85-98% | 81-96% | 87-98% | 92-103% |
EDTA plasma(n=5) | 89-101% | 79-92% | 82-93% | 83-97% |
Heparin plasma(n=5) | 82-103% | 81-91% | 85-101% | 81-90% |