Size: | Price: |
---|---|
48 T | $320.00 |
96 T | $458.00 |
Concentration (pg/mL) | OD | Corrected OD |
---|---|---|
2000.00 | 2.161 | 2.053 |
1000.00 | 1.487 | 1.379 |
500.00 | 0.975 | 0.867 |
250.00 | 0.649 | 0.541 |
125.00 | 0.493 | 0.385 |
62.50 | 0.351 | 0.243 |
31.25 | 0.315 | 0.207 |
0.00 | 0.108 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 94-107% | 100% |
EDTA plasma(n=5) | 82-95% | 88% |
Heparin plasma(n=5) | 94-105% | 98% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 86-94% | 93-102% | 87-96% | 92-105% |
EDTA plasma(n=5) | 86-96% | 93-104% | 89-97% | 85-94% |
Heparin plasma(n=5) | 85-105% | 87-98% | 91-102% | 86-102% |