Size: | Price: |
---|---|
48 T | $320.00 |
96 T | $458.00 |
Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
100.00 | 2.079 | 2.011 |
50.00 | 1.562 | 1.494 |
25.00 | 0.935 | 0.867 |
12.50 | 0.797 | 0.729 |
6.25 | 0.508 | 0.440 |
3.13 | 0.226 | 0.158 |
1.57 | 0.213 | 0.145 |
0.00 | 0.068 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 86-98% | 92% |
EDTA plasma(n=5) | 95-107% | 101% |
Heparin plasma(n=5) | 82-95% | 88% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 85-94% | 87-96% | 82-95% | 81-93% |
EDTA plasma(n=5) | 85-92% | 88-96% | 87-98% | 95-102% |
Heparin plasma(n=5) | 79-91% | 96-105% | 80-92% | 91-103% |