Products

Rabbit IL1RA(Interleukin 1 Receptor Antagonist) ELISA Kit
ELK2856
Size: Price:
48 T $412.00
96 T $588.00

Overview

Product name: Rabbit IL1RA(Interleukin 1 Receptor Antagonist) ELISA Kit
Reactivity: Rabbit
Alternative Names: IL1RN; IL1-RA; ICIL1RA; IL1ra3; IL1F3; IRAP; IL-1RA; Interleukin-1 Family Member 3; IL1 inhibitor; Anakinra
Assay Type: Sandwich
Sensitivity: 3.2 pg/mL
Standard: 500 pg/mL
Detection Range: 7.82-500 pg/mL
Sample Type: Serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay Length: 3.5h
Research Area: Cytokine;Signal transduction;Infection immunity;
Uniprot ID: P26890
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Rabbit IL1RA. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Rabbit IL1RA. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Rabbit IL1RA, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Rabbit IL1RA in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (pg/mL) OD Corrected OD
500.00 2.096 2.023
250.00 1.497 1.424
125.00 1.121 1.048
62.50 0.752 0.679
31.25 0.427 0.354
15.63 0.305 0.232
7.82 0.232 0.159
0.00 0.073 0.000

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant IL1RA and the recovery rates were calculated by comparing the measured value to the expected amount of IL1RA in samples.
Matrix Recovery range Average
serum(n=5) 82-94% 88%
EDTA plasma(n=5) 97-107% 102%
Heparin plasma(n=5) 87-99% 93%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of IL1RA and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 95-104% 79-92% 82-95% 86-97%
EDTA plasma(n=5) 86-93% 91-105% 87-98% 99-105%
Heparin plasma(n=5) 88-102% 86-92% 81-93% 89-104%