Size: | Price: |
---|---|
48 T | $412.00 |
96 T | $588.00 |
Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
100.00 | 1.849 | 1.766 |
50.00 | 1.487 | 1.404 |
25.00 | 0.976 | 0.893 |
12.50 | 0.872 | 0.789 |
6.25 | 0.541 | 0.458 |
3.13 | 0.324 | 0.241 |
1.57 | 0.211 | 0.128 |
0.00 | 0.083 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 87-99% | 93% |
EDTA plasma(n=5) | 80-93% | 86% |
Heparin plasma(n=5) | 93-107% | 100% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 87-95% | 85-98% | 82-97% | 92-101% |
EDTA plasma(n=5) | 85-98% | 81-96% | 87-98% | 92-103% |
Heparin plasma(n=5) | 81-93% | 86-98% | 91-103% | 82-95% |