Size: | Price: |
---|---|
48 T | $320.00 |
96 T | $458.00 |
Concentration (pg/mL) | OD | Corrected OD |
---|---|---|
5000.00 | 2.054 | 1.956 |
2500.00 | 1.624 | 1.526 |
1250.00 | 1.087 | 0.989 |
625.00 | 0.735 | 0.637 |
312.50 | 0.518 | 0.420 |
156.25 | 0.385 | 0.287 |
78.13 | 0.238 | 0.140 |
0.00 | 0.098 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 82-98% | 90% |
EDTA plasma(n=5) | 80-97% | 88% |
Heparin plasma(n=5) | 82-94% | 88% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 95-104% | 82-96% | 92-103% | 90-99% |
EDTA plasma(n=5) | 90-99% | 85-96% | 87-98% | 93-102% |
Heparin plasma(n=5) | 86-97% | 93-102% | 87-98% | 83-97% |