Size: | Price: |
---|---|
48 T | $320.00 |
96 T | $458.00 |
Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
100.00 | 2.126 | 2.045 |
50.00 | 1.720 | 1.639 |
25.00 | 1.235 | 1.154 |
12.50 | 0.933 | 0.852 |
6.25 | 0.614 | 0.533 |
3.13 | 0.335 | 0.254 |
1.57 | 0.215 | 0.134 |
0.00 | 0.081 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 82-95% | 88% |
EDTA plasma(n=5) | 95-107% | 101% |
Heparin plasma(n=5) | 87-99% | 93% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 86-93% | 93-101% | 86-97% | 90-101% |
EDTA plasma(n=5) | 85-94% | 87-101% | 86-97% | 88-95% |
Heparin plasma(n=5) | 88-102% | 87-102% | 85-96% | 87-101% |