Products

Human IL29(Interleukin 29) ELISA Kit
ELK1784
Size: Price:
48 T $320.00
96 T $458.00

Overview

Product name: Human IL29(Interleukin 29) ELISA Kit
Reactivity: Human
Alternative Names: IL-29;FNL1; Interferon,Lambda 1; Cytokine Zcyto21
Assay Type: Sandwich
Sensitivity: 2.7 pg/mL
Standard: 500 pg/mL
Detection Range: 7.82-500 pg/mL
Sample Type: Serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay Length: 3.5h
Research Area: Cytokine;Infection immunity;
Uniprot ID: Q8IU54
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human IL29. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human IL29. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Human IL29, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human IL29 in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (pg/mL) OD Corrected OD
500.00 2.035 1.946
250.00 1.717 1.628
125.00 1.035 0.946
62.50 0.782 0.693
31.25 0.469 0.380
15.63 0.315 0.226
7.82 0.206 0.117
0.00 0.089 0.000

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant IL29 and the recovery rates were calculated by comparing the measured value to the expected amount of IL29 in samples.
Matrix Recovery range Average
serum(n=5) 96-107% 101%
EDTA plasma(n=5) 90-105% 97%
Heparin plasma(n=5) 85-97% 91%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of IL29 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 86-105% 84-98% 95-102% 88-106%
EDTA plasma(n=5) 87-94% 91-104% 95-107% 82-96%
Heparin plasma(n=5) 81-94% 93-107% 93-102% 91-105%