Products

Testo(Testosterone) ELISA Kit
ELK1332
Size: Price:
48 T $265.00
96 T $376.00

OverviewPublications

Product name: Testo(Testosterone) ELISA Kit
Reactivity: General
Alternative Names: T; Testesteron; Testeronun
Assay Type: Competitive Inhibition
Sensitivity: 17.7 pg/mL
Standard: 4000 pg/mL
Detection Range: 62.5-4000 pg/mL
Sample Type: serum, plasma, other biological fluids
Assay Length: 2h
Research Area: Endocrinology;Reproductive science;Hormone metabolism;
Test principle: This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with Testo protein. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Testo. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Testo in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (pg/mL) OD Corrected OD
4000.00 0.183
2000.00 0.332
1000.00 0.567
500.00 0.823
250.00 1.157
125.00 1.537
62.50 1.715
0.00 2.213

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant Testo and the recovery rates were calculated by comparing the measured value to the expected amount of Testo in samples.
Matrix Recovery range Average
serum(n=5) 90-103% 96%
EDTA plasma(n=5) 81-96% 88%
Heparin plasma(n=5) 78-94% 86%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Testo and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 85-94% 90-99% 95-102% 82-90%
EDTA plasma(n=5) 85-96% 87-101% 89-102% 85-98%
Heparin plasma(n=5) 86-102% 87-96% 89-98% 83-103%