Products

Human CLU(Clusterin) ELISA Kit
ELK1079
Size: Price:
48 T $265.00
96 T $376.00

Overview

Product name: Human CLU(Clusterin) ELISA Kit
Reactivity: Human
Alternative Names: CLU; TRPM2; KUB1; CLI; APOJ; SGP2; AAG4; APO-J; SP-40; Apolipoprotein J; Complement Lysis Inhibitor; 40,Sulfated Glycoprotein 2; Testosterone-Repressed Prostate Message 2
Assay Type: Sandwich
Sensitivity: 5.1 ng/mL
Standard: 800 ng/mL
Detection Range: 12.5-800 ng/mL
Sample Type: serum, plasma, urine, saliva, tissue homogenates, cerebrospinal fluid, cell culture supernates and other biological fluids
Assay Length: 3.5h
Research Area: Apoptosis;Tumor immunity;Infection immunity;
Uniprot ID: O75153
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human CLU. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human CLU. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Human CLU, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human CLU in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (ng/mL) OD Corrected OD
800.00 2.219 2.111
400.00 1.674 1.566
200.00 1.354 1.246
100.00 0.896 0.788
50.00 0.498 0.390
25.00 0.364 0.256
12.50 0.225 0.117
0.00 0.108 0.000

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant CLU and the recovery rates were calculated by comparing the measured value to the expected amount of CLU in samples.
Matrix Recovery range Average
serum(n=5) 88-102% 95%
EDTA plasma(n=5) 82-97% 89%
Heparin plasma(n=5) 90-105% 97%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of CLU and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 93-102% 78-92% 87-98% 80-93%
EDTA plasma(n=5) 88-97% 86-92% 95-103% 98-106%
Heparin plasma(n=5) 95-102% 87-101% 82-91% 85-97%