Products

Human ALB(Albumin) ELISA Kit
ELK1071
Size: Price:
48 T $265.00
96 T $376.00

Overview

Product name: Human ALB(Albumin) ELISA Kit
Reactivity: Human
Alternative Names: Serum albumin
Assay Type: Competitive Inhibition
Sensitivity: 89.6 ng/mL
Standard: 20000 ng/mL
Detection Range: 312.5-20000 ng/mL
Sample Type: serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay Length: 2h
Research Area: Tumor immunity;Infection immunity;Hematology;Hepatology;Nutrition metabolism;
Uniprot ID: P02768
Test principle: This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with Human ALB protein. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human ALB. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human ALB in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (ng/mL) OD Corrected OD
20000.00 0.179
10000.00 0.234
5000.00 0.342
2500.00 0.531
1250.00 0.977
625.00 1.464
312.50 1.786
0.00 2.475

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant ALB and the recovery rates were calculated by comparing the measured value to the expected amount of ALB in samples.
Matrix Recovery range Average
serum(n=5) 81-95% 88%
EDTA plasma(n=5) 95-103% 99%
Heparin plasma(n=5) 92-102% 97%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of ALB and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 95-106% 85-101% 87-102% 89-95%
EDTA plasma(n=5) 95-102% 93-101% 81-89% 85-99%
Heparin plasma(n=5) 85-95% 79-96% 89-101% 85-97%