Size: | Price: |
---|---|
48 T | $320.00 |
96 T | $458.00 |
Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
10.00 | 2.252 | 2.136 |
5.00 | 1.642 | 1.526 |
2.50 | 1.243 | 1.127 |
1.25 | 0.876 | 0.760 |
0.63 | 0.596 | 0.480 |
0.32 | 0.384 | 0.268 |
0.16 | 0.268 | 0.152 |
0.00 | 0.116 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 78-95% | 86% |
EDTA plasma(n=5) | 80-96% | 88% |
Heparin plasma(n=5) | 81-95% | 88% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 93-102% | 86-92% | 96-105% | 89-97% |
EDTA plasma(n=5) | 90-103% | 97-106% | 82-90% | 92-101% |
Heparin plasma(n=5) | 85-92% | 82-96% | 87-105% | 87-92% |