Products

Goat GC(Glucagon) ELISA Kit
ELK0877
Size: Price:
48 T $265.00
96 T $376.00

Overview

Product name: Goat GC(Glucagon) ELISA Kit
Reactivity: Goat
Alternative Names: GCG; GLP1; GLP2; GRPP; Glicentin-Related Polypeptide; Glucagen; Oxyntomodulin; Incretin hormone
Assay Type: Competitive Inhibition
Sensitivity: 12.3 pg/mL
Standard: 2000 pg/mL
Detection Range: 31.25-2000 pg/mL
Sample Type: serum, plasma and other biological fluids
Assay Length: 2h
Research Area: Endocrinology;
Test principle: This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with Goat GC protein. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Goat GC. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Goat GC in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (pg/mL) OD Corrected OD
2000.00 0.195
1000.00 0.375
500.00 0.524
250.00 0.823
125.00 1.163
62.50 1.557
31.25 1.755
0.00 2.211

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant GC and the recovery rates were calculated by comparing the measured value to the expected amount of GC in samples.
Matrix Recovery range Average
serum(n=5) 80-97% 88%
EDTA plasma(n=5) 82-95% 88%
Heparin plasma(n=5) 80-95% 87%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of GC and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 89-104% 97-106% 82-91% 85-96%
EDTA plasma(n=5) 95-102% 96-104% 87-98% 95-107%
Heparin plasma(n=5) 92-103% 81-92% 78-87% 91-99%