Products

Rat Glu(Glutamic Acid) ELISA Kit
ELK0812
Size: Price:
48 T $265.00
96 T $376.00

Overview

Product name: Rat Glu(Glutamic Acid) ELISA Kit
Reactivity: Rat
Alternative Names: Glutamate
Assay Type: Competitive Inhibition
Sensitivity: 48.2 ng/mL
Standard: 10000 ng/mL
Detection Range: 156.25-10000 ng/mL
Sample Type: biological fluids
Assay Length: 2h
Research Area: Enzyme & Kinase;Endocrinology;Neuro science;
Test principle: This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with Rat Glu protein. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Rat Glu. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Rat Glu in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (ng/mL) OD Corrected OD
10000.00 0.199
5000.00 0.397
2500.00 0.611
1250.00 0.823
625.00 1.253
312.50 1.596
156.25 1.863
0.00 2.115

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant Glu and the recovery rates were calculated by comparing the measured value to the expected amount of Glu in samples.
Matrix Recovery range Average
serum(n=5) 87-99% 93%
EDTA plasma(n=5) 80-96% 88%
Heparin plasma(n=5) 94-105% 98%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Glu and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 95-101% 88-102% 79-92% 85-96%
EDTA plasma(n=5) 87-96% 82-93% 89-97% 85-93%
Heparin plasma(n=5) 95-106% 87-96% 82-90% 92-101%