Products

Rat RAB3A(Ras-Related Protein Rab-3A) ELISA Kit
ELK0557
Size: Price:
48 T $320.00
96 T $458.00

Overview

Product name: Rat RAB3A(Ras-Related Protein Rab-3A) ELISA Kit
Reactivity: Rat
Alternative Names: Ras-related protein Rab-3A
Assay Type: Sandwich
Sensitivity: 31.2 pg/mL
Standard: 5000 pg/mL
Detection Range: 78.13-5000 pg/mL
Sample Type: tissue homogenates, cell lysates and other biological fluids
Assay Length: 3.5h
Research Area: Tumor immunity;
Uniprot ID: P63012
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Rat RAB3A. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Rat RAB3A. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Rat RAB3A, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Rat RAB3A in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (pg/mL) OD Corrected OD
5000.00 2.266 2.179
2500.00 1.628 1.541
1250.00 1.226 1.139
625.00 0.876 0.789
312.50 0.593 0.506
156.25 0.385 0.298
78.13 0.213 0.126
0.00 0.087 0.000

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant RAB3A and the recovery rates were calculated by comparing the measured value to the expected amount of RAB3A in samples.
Matrix Recovery range Average
serum(n=5) 78-90% 84%
EDTA plasma(n=5) 85-97% 91%
Heparin plasma(n=5) 80-93% 86%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of RAB3A and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 89-99% 88-104% 84-98% 95-105%
EDTA plasma(n=5) 88-97% 96-103% 86-92% 89-97%
Heparin plasma(n=5) 88-97% 93-103% 94-106% 91-101%